All screens can be selected for
display in English or Spanish.
Different languages such as
French or German may also be
substituted.
The simple keypad and logical
menu structure make this
transmitter easy to use. Menu
screens guide you through setup,
operation, calibration, and
test/maintenance functions.
For security, use the passcode
feature to restrict configuration
and calibration settings to only
authorized personnel.
The isolated 4-20 mA analog
output can represent the
measured pH or temperature or
ORP. During calibration, the
analog output is automatically
held at the last measured value
and, upon completion, returned to
its active state.
|
|
The PRO-P3 transmitter can be
used with any GLI Differential
Technique pH or ORP sensor, or
any conventional combination pH
or ORP electrode.
Automatic temperature
compensation is provided when
using NTC 300 ohm thermistor,
Pt 1000 RTD or Pt 100 RTD
temperature elements. For
applications requiring fixed
temperature compensation, the
PRO-P3 can be manually set to a
desired temperature.
Built-in diagnostics continuously
test transmitter and sensor
operation.
PRO-series transmitters can be
packaged or configured to
accommodate OEM-specific
needs. |