Hach ldo Dissolved Oxygen Probe


 
ldo®

 

 

Features & Benefits

Method of Detection.
The HACH LDO sensor is coated with a luminescent material. Blue light from an LED is transmitted to the sensor surface. The blue light excites the luminescent material. As the material relaxes it emits red light. The time for the red light to be emitted is measured. Between the flashes of blue light, a red LED is flashed on the sensor and used as an internal reference.
Increased oxygen in the sample decreases the time it takes for the red light to be emitted. The time measurements correlate to the oxygen concentration.

Accuracy & Speed.
The sensor is unaffected by pH swings, hydrogen sulfide, wastewater chemicals, heavy metals, or organic build-up on the sensor. Warm-up time is unnecessary so the analyzer can start measuring within 30 seconds of when it’s turned on.

Simple Installation & Maintenance.
The HACH LDO probe can be installed with a choice of pole or ball-float mounting kits. The probe has only one inexpensive replacement part—the sensor cap that is simple to replace. The sensor cap is warranted for one year. The probe is warranted for three years.
An internal standard calibrates the instrument with every reading. Factory calibration and membrane replacement are unnecessary. Periodic wiping of the sensor with a wet rag is all that is required for continuous measurement.

For Pricing Call Toll Free - 877-215-9090

 

 

Applications
„Aeration Tanks.
Adequate dissolved oxygen levels in aeration basins are required for microorganisms to remain viable. The HACH LDO probe can be used with a variable frequency drive or PLC control system to control the amount of oxygen being injected into the tank.

Nitrification & Denitrification Tanks

Anaerobic & Aerobic Digesters

NPDES Permit Monitoring
Up & downstream from plant effluent and plant outfall.

 

Full Featured “Plug and Play” SC-100 Digital Controller
There’s no complicated wiring or set up procedures with the Hach sc100 controller. Just plug in any Hach digital sensor and it’s ready to use—it’s “plug and play.”

One or two sensors—Use the sc100 Digital Controller to receive data from up to two Hach digital sensors in any combination.

Communications—Multiple alarm/control schemes are available using three relays and two PID control outputs. Communications use analog 4-20 mA and digital MODBUS®/RS485, MODBUS®/RS232 protocols. (Other digital protocols are available. Contact your Hach representative for details.) Every sc100 controller is equipped with wireless communication through an infrared port.

Data logger—A built-in data logger collects measurement data, calibration, verification points, and alarm history for up to 6 months.

 

  • Specifications
  • Dimensions
  • Engineering
  • Ordering
  • Installation Examples

lodspec

 

 

ldo dimen

 

ldoengin

ldoorderldo2

ldoinstall

 

 
Proud Members of the

Now Accepting

cc

Industrial Instrumentation Services, Inc.
1400 Rahway Avenue, Suite 4 • Avenel, NJ • 07001-2214
Phone: 732-815-9090 • Fax: 732-815-9092
Toll Free: 877-215-9090