hm70 hand-held humidity and temperature meter for spot-checking applications


 
Kaye Validator 2000
The Vaisala HUMICAP® Hand-Held Humidity and Temperature Meter HM70 is a high- performance, portable humidity reference that is exciting to use

 

 

  • Proven Vaisala HUMICAP® Sensor technology
  • 3 probes, temperature measurement ranges between -70 and +180 °C
  • RH measurement range 0…100 % RH
  • Multipurpose operation; dewpoint and CO2 probes can also be connected
  • Displays various humidity parameters
  • Sensor preheat and chemical purge options for demanding conditions
  • Data can be logged, and transferred to a PC via MI70 Link software
  • Rugged and reliable - even in extreme conditions
  • NIST traceable (certificate included)
  • Multilingual user interface
  • Shows measurement trends graphically

For Pricing Call Toll Free - 877-215-9090

 

 

Three probes to choose from
The HMP75 is a general purpose probe for humidity measurements between -20 and +60 °C. The HMP76 is a long, stainless steel probe for humidity measurements between -50 and +120 °C. It is especially suitable for spotchecking in ducts. The HMP77 probe has a small sensor head at the end of a 5-meter cable. The probe is ideal for difficult-to-reach areas and for on-site calibration of Vaisala's process transmitters. The measurement range is for temperatures up to +180 °C. One or two probes can be connected simultaneously. In addition, the HM70 supports the use of Vaisala's dewpoint, carbon dioxide and moisture in oil probes, allowing measurements in multiparameter applications.

Easy-to-use user interface and three-
variable display

The HM70 features a multilingual, menu-based user interface, and a clear graphical LCD display. Overall three measurement parameters can be displayed and logged into the meter's memory at the same time. An analog output option is also available.

 

Proven HUMICAP® technology
The HM70 incorporates Vaisala HUMICAP® Sensor, which provides reliable and high performance humidity measurement with excellent long-term stability. Options improve performance in demanding applications The chemical purge option maintains measurement accuracy in environments with high concentrations of chemicals. The sensor preheat option reduces measurement delays by keeping the sensor dry when the probe is inserted into hot and humid processes.

Calibration
The meter is calibrated at the factory against traceable standards, and delivered with a calibration certificate. Also, the HM70 is available with accredited calibration. The HM70 probes can be sent to a Vaisala Service Center for a traceable recalibration.

MI170 Link
The optional MI70 Link Windows® software is a practical tool for transferring logged data and real time measurement data from the HM70 to a PC.

 

 

  • Technical Data
  • Probe
  • MI70 Indicator
  • HM70
  • Dimensions

Relative humidity measurement

Measurement range
Accuracy at +20 °C (68 °F)

Sensor

0 ... 100 % RH
± 1.0 % RH (0 ... 90 % RH)
± 1.7 % RH (90 ... 100 % RH)
Vaisala HUMICAP® 180

Temperature measurement

Measurement range
HMP75
HMP76
short time
HMP77
Accuracy at +20 °C (+68 °F)
Sensor


-20 ... +60 °C (-4 … +140 °F)
-50 ... +120 °C (-58 … +248 °F)
-50 ... +180 °C (-58 … +356 °F)
-70 ... +180 °C (-94 … +356 °F)
± 0.2 °C (0.36 °F)
Pt100 1EC751 1/3 class B

Operating environment

Temperature (MI70 measurement indicator)
Humidity


-10 ... +40 °C (+14 ... +104 °F)
non-condensing

Inputs and outputs

Power consumption

Analog output
Menu languages

Rechargeable NiMH battery pack with AC power or 4 x AA alkaline batteries, type IEC LR6
0 ... 1 VDC
English, French, Spanish, German, Swedish, Finnish, Russian, Japanese

Mechanics

Weight
Housing classification

400 g
IP54

General

Operating temperature range for electronics -40...+60 °C
(-40...+140 °F)
Housing classification IP65 (NEMA 4)
Housing material ABS/PC blend
Probe material Stainless steel (AISI316L)
Cable length between probe and indicator 1.9 m
Weight
HMP75 250 g
HMP76 350 g
HMP77 500 g HUMICAP® is a registered trademark of Vaisala.

Indicator general
Menu languages: English, French, Spanish, German, Swedish, Finnish
Display LCD with backlight, graphic trend display of any parameter, character height up to 16 mm
Probe inputs 1 or 2
Power supply Rechargeable NiMH battery pack with AC-POWER
or 4xAA-size alkalines, type IEC LR6
Analog output 0...1 VDC
Output resolution 0.6 mV
Data interface RS-232 (accessible only with MI70 Link software)
Data logging capacity 2700 points
Alarm audible alarm function
Operating temperature range -10...+40 °C (+14...+104 °F
Operating humidity range non-condensing
Housing classification IP54
Weight 400 g
Battery operation time
Continuous use 48 h typical at +20 °C (+68 F)
Datalogging use up to 30 days
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Portable Equipment
MI70 indicator + probe = HM70
Accessories
Carrying cases
For MI70 and HMP75/77 probe MI70CASE
For MI70 and HMP75/7 probe MI70CASE2
Transmitter & connection cables for
HMP230 Series 27159ZZ
HMT330 & HMT320 Series 211339
HM60/70/140 Series HMA6070
HM20/30/130 Series HMA2030
Analog output cable 27168ZZ
Windows® software, includes PC connection cable MI70LINK

Dimensions in mm (inches)

vaisaladimen

 

 

 

 
Proud Members of the

Now Accepting

cc

Industrial Instrumentation Services, Inc.
1400 Rahway Avenue, Suite 4 • Avenel, NJ • 07001-2214
Phone: 732-815-9090 • Fax: 732-815-9092
Toll Free: 877-215-9090

© Copyright 2006 INDUSTRIAL INSTRUMENTATION SERVICES, INC. (IIS) All Rights Reserved